ARC 400

ARC electrosurgery unit

Поделиться

Excellent results depend on the right technology. In
addition to the innovative socket design that allows
up to two monopolar and three bipolar instruments
to be attached simultaneously and is compatible
with all popular standard connections, there are
numerous hardware and software components that
can also directly affect performance in the OR.
– ARC control arc regulation for reproducible
monopolar and bipolar cutting results
– Up to 9 effects per mode
– EASY neutral electrode monitoring with contact
monitor, baby mode with automatic power limitation
– Current leakage monitoring
– Short circuit detection
– ISSys permanent system self-testing
– CCS permanent initial cutting support
– Configurable socket design
– Information area for device messages
– Operating instructions available directly on the
device
– CUT modes: e. g. Standard, Dry, Cardiac,
GastroCut, Loop and Knife, MetraLOOP, Laparoscopy
– Micro-regulations and power forms for Plastic and
Neurosurgery down to 0.1 W
– COAG Modes: Moderate, Forced, Spray, Cardiac
– Autostart function

Monopolar simultaneous coagulation
permits the simultaneous COAG activation of two
monopolar handles. This mode is particularly ideal
for coagulation and preparation, with the selected
power setting being available for both handles.
Applications include mastectomies in gynaecology,
bypasses in cardiac surgery or multiple traumas in
surgery.

Technical data at a glance ARC 400 ARC PLUS
Mains voltage 100 – 120 V / 220 – 240 V ± 10 % 100-240 V ± 10%
Mains frequency 50 / 60 Hz 50 / 60 Hz
Mains power Max. 10 A/5A Max. 0.6 A
Mains fuse T 10 A/5 A T 1 A
Power consumption in standby mode 40 W 6 W
Power consumption at max. 550 W 35 W
Width x Height x Depth 430 x 180 x 475 mm 430 x 90 x 490 mm
Weight 12.5 kg 7.7 kg
Classification under EC Directive 93/42/EEC II b II a
Protection class according to EN 60601-1 1 1
Type according to EN 60 601-1 CF CF
CE marking CE0123 CEO123
Item no. 900-400 900-001
Bipolar resection option 900-395  
LIGATION option 900-396  
Bipolar SimCOAG option 900-399  
Maximum CUT power 400 W  
Maximum COAG power 350 W  
Mains frequency 350 kHz / 1 MHz